The first objective was achieved by assessing the effectiveness of the containment provided by the left chamber of the test isolator (Figure 2) to verify there were no inherent weak points with the isolator. The task used to challenge the containment of the left chamber involved transferring a small quantity (500 g) of surrogate powder (acetaminophen), from one container to another, using a spoon. Proximate samples were placed around the outside of the left chamber at potential weak points, for example at glove ports and isolator seals. Proximate samples were also placed inside the right chamber adjacent to the hinged stainless-steel plate that housed the DPTE Alpha port. The assessment consisted of three test runs; each test run was identical in nature except for the size of the DPTE Alpha port apertures fitted to the right side of the left chamber. A different port aperture was used for each of the three test runs to verify that changing of the DPTE diameter would not affect the containment performance of the isolator. The likelihood of surrogate emissions migrating from the left chamber through the hinged stainless-steel plate and DPTE Alpha port was concluded to be extremely low. During the preliminary containment verification of the proposed isolator, twelve air samples were taken from positions adjacent to the stainlesssteel plate and DPTE Alpha port and only 3 samples returned values at detectable concentrations, the highest recorded value was 3 ng/m3. PRELIMINARY WORK: CLEANING VALIDATION Assessing the potential for airborne and surface contamination to occur when changing the size of the DPTE Alpha port aperture (objective two) was fundamental in allowing the comparison of containment performance data between each of the DPTE Alpha port apertures. When reviewing the proposed isolator setup it was apparent that changing the aperture of the DPTE Alpha port (ready for the next containment performance assessment) could lead to an unwanted release of acetaminophen from the potentially contaminated and open left chamber and the surface of the stainless-steel plate which housed the DPTE Alpha port (Figure 3). The methodology developed for limiting the impact of this potential contamination when changing the DPTE Alpha port aperture focused on adequately cleaning the internal surfaces of the left chamber and preventing any transfer of surface contamination to the right chamber. This was achieved by developing a robust cleaning method and verifying the effectiveness of the cleaning by quantitative air and surface sampling. The following operations occurred during the cleaning validation: • After the cleaning of the left chamber of the isolator, the isolator was left running overnight (at least 12 hours) at -100 pa. • Surface samples from 5 positions were then taken inside the left chamber. • The DPTE Alpha port was then replaced with a different sized DPTE alpha port. • The right chamber of the isolator was then cleaned. • Surface samples from 4 positions were taken inside the right chamber. • Two air samples were taken inside the right chamber close to the DPTE Alpha port. Chimica Oggi - Chemistry Today 24 vol. 42(2) 2024 Figure 2. Test isolator used for containment performance assessment of DPTE Alpha port. Figure 3. Stainless-steel plate housing the DPTE Alpha port.
RkJQdWJsaXNoZXIy MTY5Njg=